3B SCIENTIFIC
®
PHYSICS
1
Digital Multimeter E 1006809
Instruction sheet
05/12 ALF
1 Measuring probe
1a Finger guards
2 Measurement socket “20 A“for
current measurement 20 A
range (plus)
3 Measurement socket “A” for
current measurement for up to
2 A (plus)
4 Measurement socket “COM“
(minus)
5 Measurement socket “V/“ for
voltage and resistance meas-
urement (plus)
6 Transistor test socket
7 Measurement range dial
8 Button Hold function
9 On/Off switch
10 Display
11 Unlock button
!
1. Safety instructions
The Digital Multimeter E conforms to the safety
requirements for electrical equipment for meas-
urement, control and laboratory use in DIN EN
61010 part 1. Safe operation of the apparatus is
guaranteed with correct handling. However,
safety is not guaranteed if the apparatus is han-
dled improperly or carelessly.
Read this manual carefully before using the
digital multimeter and follow the instructions!
Before using the meter, check the case and
test leads for any damage. In the event of
any malfunction/operational defect or visible
damage, do not use the meter. Pay particu-
lar attention to the insulation surrounding the
measurement sockets.
Use with caution when working above 30 V
ACrms, or 60 V DC. Such voltages pose a
shock hazard.
The limit of the measurement range must
not be exceeded. If the values of the meas-
urand are unknown, always switch from a
higher measurement range to a lower one.
When measuring current, turn off circuit
power before connecting the meter in the
circuit.
Connect the common test lead before you
connect the live test lead. When you discon-
nect test leads, disconnect the live test lead
first.
When using the measuring probes, keep
your fingers behind the finger guards on the
probes.
Do not operate the meter around explosive
gas, vapour, or dust.
Do not conduct measurements in a humid
environment. Work area, hands, shoes and
floor must be dry.
To avoid false readings, which could lead to
possible electric shock or personal injury,
replace the batteries as soon as the low bat-
tery indicator (
+ -
) appears.
Before the case is opened, the meter has to
be switched off and the leads must be dis-
connected from the meter.
Never use the meter when the case is open.
When disposing empty batteries follow the
local regulations. Never dispose of them in
the regular household garbage.
Page view 0
1 2 3 4

Summary of Contents

Page 1 - PHYSICS

3B SCIENTIFIC® PHYSICS 1 Digital Multimeter E 1006809 Instruction sheet 05/12 ALF 1 Measuring probe 1a Finger guards 2 Measurement socket

Page 2

2 2. Symbol legend ! Read instruction sheet Dangerous voltages V DC voltage A DC current V AC voltage A AC current Diode and continuity

Page 3

3 A Measuring range Accuracy 20 µA, 200 µA, 2 mA 20 mA ±2.0 % ± 3 digits 200 mA ±2.0 % ± 5 digits 2 A, 20 A ±2.5 % ± 10 digits AC current Measur

Page 4

Elwe Didactic GmbH ▪ Steinfelsstr. 5 ▪ 08248 Klingenthal ▪ Germany ▪ www.elwedidactic.com 3B Scientific GmbH ▪ Rudorffweg 8 ▪ 21031 Hamburg ▪ Germany